Configuring for serial output

The Baud rate settings set out below must be matched in the Nexus Cloud Platform

Setup > Despatch > Serial Settings

Enable Serial Port

Baud 9600

Data Bits/Parity 8 Data Bits, No Parity

image.png

Address - Add the address of the callpoints in the fields. Selecting "0" To "0" will cover all callpoints

Event - As a minimum select 128 -129 - You can select every event so we capture all events from the nursecall.

Despatch Type - Serial 

Repeat - Add the amount of seconds to resend the alert if not reset at call point

Auto Cancel Repeats > Enable

image.png

Use a serial program such as putty to test the connection and that you are getting the output for all events specified above.

 

image.png